Product Details
- Available Certified ISO Grade Designs: ISO 7, ISO 8
- Six-channel laser diode portable airborne particle counter for ISO 1-8 cleanrooms
- Cleanroom certification ISO 21501-4 Compliant
- Pass/fail report wizard for ISO-14644-1, EU GMP Annex 1, and FS-209E report formats
- Light Source: Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF)
- Connect Solair model to a manifold for real-time cleanroom particle contamination monitoring
- Includes: Operating manual with CD, LMS Express transfer software download instructions, isokinetic sample probe, tripod, tubing, purge filter, power cables, spare fuse, printer, printer paper
- Meets ISO 21501-4 calibration
- Applications: cleanroom certification, troubleshooting, routine testing, compressed gas testing, HEPA filter scanning, HEPA filter challenge testing, environmental testing
- Light Source: Extreme Life Laser Diode Technology with >20-year Mean Time To Failure (MTTF)
- Zero Count Level<1 count / 5 minutes (per JIS B 9921) on 3100, 3350, 5100 and Boulder models
- Flow Rate: 100 LPM
- Calibration Meets ISO 21501-4 calibration using NIST traceable PSL spheres, DMA and condensation Particle Counter
- Security: Integrated alarm buzzer and external alarm connector, alarms on counts, low battery, sensor failure
- Sample Output: Internally filtered to HEPA standards (>99.97% @ 0.3μm)
- Features: stainless steel enclosure; icon-driven 5.7 - inch (14.47 cm) color TFT touch screen display; USB, Ethernet, RS-485, Secure Data Transfer to and Storage on USB Flash Drive; automatic, manual, beep, concentration, cumulative/differential count modes; rotating buffer (includes particle and environmental data, location and time); Languages English (U.S.), French, German, Italian, Spanish, Chinese, Russian, Japanese, Korean
- Vacuum Source: Internal closed-loop controlled pump. Allows the use of 3/8" ID sample tubing of up to 12 meters (40 feet) @ 50 & 100L
- Optional Environmental Sensors(up to 4) temperature/relative humidity, air velocity, differential pressure or others
- Supporting Software (sold separately): LMS Express, LMS Express RT, LMS Pharma
- Optional channels available on all Solair Airborne Particle Counter models
Specs
• Detection Channels: 6 Channels
• Detection Range: 0.3 - 25 µm
• Counting Efficiency: 50% @ 0.3 µm, 100% > 0.45 µm
• Manufacturer SKU: Solair 3350
• Unit of Measure: EA of 1
• Electrical Specifications: 100/240 V, 300 W, 50/60 Hz
• Weight, Product: 23.05 lb
• Dimensions, Product: 8" W x 14.65" D x 10.4" H