Product Details
- Available Certified ISO Grade Designs: ISO 5, ISO 6, ISO 7, ISO 8
- Construction: Interior coved corners and radius bends
- 180-grit pharmaceutical grade (#4) finish facilitates cleaning and decontamination
- Exhaust System Requirements: 50 cfm @ 2" H2O (8.5 M3/hr. @ 500 Pa)
- Complies with standard EN 12469-2000
- Airlock doors include PLC controlled, double-door interlocking mechanism
- Interior airlock door opening: 12" W x 13" H (305 x 330 mm)
- Front facing airlock door opening: 12" W x 10" H (305 x 250 mm)
- Gloveports: Sealed into the window at an ergonomic "V" angle
- Double groove gloveports allow glove changes without breaking containment
- Fitted with one set of 815 mm long Hypalon(R) (CSM) gloves, size 9.75 ambidextrous
- Viewing Window: angled and sealed to glovebox with airtight gaskets
- Inlet and exhaust HEPA filters: Rated at 99.99% efficient at a 0.3μm particle size;
- HEPA filters can be individually aerosol tested, sealed to glovebox with stainless steel frame
- 76mm diameter inlet and exhaust ducts with sanitary shut off butterfly valves
- Removable bolt on side panel for setup of equipment prior to use. (After decontamination)
- All interior surfaces reachable through gloveports for surface wipe-down without breaking containment
- All welded tubular stainless steel stand
- Safe sample entry with double door airlock that can be mounted on left or right side (specify upon ordering)
- Includes passive HEPA purge
- Easy to access panel features: digital and mechanical pressure gauges
- Low-pressure alarm with audible and visual functions
- Separate main and light on/off power switches and circuit breaker reset
- Exterior mounted non-glare fluorescent light
- Exhaust blower mounted on stand below glovebox
- Two sealed electrical receptacles provided in work area
- Product categorized by the Export Administration Regulations (EAR) under Export Control Classification Number (ECCN) 2B352: Biological Processing Equipment. Export of this product outside of the United States may require a license to be obtained by the exporter, unless the client holds an active license exemption. Any fees associated with obtaining an export license will be borne by the customer.
Specs
• Material Type: Stainless Steel
• Application: Sterile Drug Compounding (USP 797)
• Material: 304 Stainless Steel
• Number of Glove Ports: 3
• Window Material: Acrylic
• Maximum Working Height: 30"
• Chamber Width: 22"
• Chamber Depth: 22"
• Chamber Height: 30"
• Unit of Measure: EA
• Marks & Listings: CE, UL, USP 797
• Electrical Specifications: 120 V, 15 A, 1,500 W, 60 Hz
• Dimensions, Product: 76" W x 25" D x 38" H
• Weight, Shipping: 760 lb
• Dimensions, Shipping: 82" W x 31" D x 42" H
• Package Type: Palletized