FDL 115 Options and Accessories
Additional Features of FDL Series Safety Drying Chamber: More Information
- Easy-to-change fresh air filter cartridges (replaceable) ensures constant fresh air quality
- Symmetrical air flow with electronic monitoring system and automatic deactivation of heating
- APT.line™ preheating technology provides optimal temperature uniformity for faster recovery times
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- Controller: time-segment and real-time programming
- 60-mm-thick insulation, keeping heat dissipation to a minimum
Options and Accessories for FDL 115 Series for Specimens Containing Solvents
The FDL 115 Safety Drying Chamber is available with a range of accessories and options to satisfy a variety of applications in norm-compliant drying applications containing solvents.
- Access port with silicone plug (for external measuring equipment; large selection of diameters and positions (left, right, or top)
- Calibration certificate, expanded
- Calibration certificate, temperature
- Cart with casters and locking brakes
- Coil coating door flap ideal for coil coating tests
- Door gasket made of high temperature resistant silicone for operational temperatures above 200 °C
- Door lock
- Measurement of air exchange rate
- pH-neutral detergent
- Rack
- Rubber pads
- Shelf (perforated, stainless steel
- Software (Version 2 to 3, GLP, Basic, Standard and GLP DEMO editions)
- Specimen temperature display
- Services: Access our installation, maintenance, calibration, validation, and warranty services by taking out a maintenance service agreement
Model | FDL 115 |
Interior volume | 4.1 cu. ft. (116 L) |
Outer dimensions (W x D x H) | 32.7 x 27 x 31.7 in. (831 x 686 x 806 mm) |
Internal Dimensions(W x D x H) | 23.64 x 17.14 x 17.14 in. (601 x 435 x 435 mm) |
Electrical data | 230V, 50/60Hz |
Net weight of the unit (empty) | 198 lbs. (89.8 kg) |
Temperature range 10 °C above ambient temperature to [°C] | 300 |
Wall clearance back | 6.3 in. (160 mm) |
Wall clearance sidewise | 3.94 in. (100 mm) |
Temperature variation at 150 °C [± K] | 2.5 |
Temperature fluctuation [± K] | 0.8 |
Heating-up time to 150 °C [min] | 20 |
Recovery time after 30 seconds door open at 150 °C [min] | 12 |
Air circulation (approx.) [x/min] | 20 |
Volumetric flow rate of exhaust air according to EN 1539 at 50 °C [ca. L/min] | 400 |
Highest permitted solvent quantity (at T-180 °C, M-100g/mol, U-40g/m3, K=0,5) [g] | 6.65 |
Nominal power [kW] | 2.9 |
Phase (Nominal voltage) | 1~ |
Energy consumption at 150 °C [Wh/h] | 1200 |
Sound-pressure level [dB(A)] | 57 |
Permitted load | 132 lbs. (59.9 kg) |
Load per rack | 66 lbs. (29.9 kg) |
Number of shelves (std./max.) | 2/5 |