Learn More: Extraction and Sample Parameters
More Information
- Automatic setting of elution volumes from 20 and 500 μl
- Prevents cross contamination
- Automated transfer of eluates to capped tubes for storage
- Processes forensic materials, processed foods and other complex samples
- Reliable protection for the dispensing unit with 1 ml pipette tips with integrated aerosol filter
Basis | Magnetic or paramagnetic particles |
Duration of extraction | - Externe Lysis: < 45 min (ohne Lyse)
- Interne Lyse: < 75 min (inkl. Lyse)
|
Ø Yield | - Depending on type and amount of sample
- Tissue (20 mg): up to 50 μg
- Whole blood samples (200 μl): up to 10 μg
- Plant (100 mg): up to 60 μg
|
Elution volume | 20 μl up to 500 μl (in steps of 10 μl) |
Field of application | DNA and RNA extraction |
Number of samples | Up to 16 samples in parallel, incl. single sample preparation |
Sample amount | - Depending on the type of sample
- Up to 10 ml or 180 mg respectively
|
Tempering | - Up to 70 °C inside the sample
- Support of lysis, drying and elution
|
Control | Stand-alone control |
Display | 10“ tablet PC, color, touch, WIN 8.1 |
Features | - Sample-based wizard
- Automatic pipetting of chosen elution volume
- Automatic transfer of eluates into storage tubes with lid
- Pre-installed extraction and decontamination protocols
- Video sequences and display of residual time for ideal overview of the run
- User management with 3 user levels
|
Pipetting head / Channels | Dosing unit with 16 channels |
Tips | 1 ml tip with aerosol filter |
Working volume | Up to 1000 μl |
Functions | Pipetting and piercing function of the sealed reagent plastic |
Cleaning / Decontamination | - Big front door for easy access and wipe decontamination of the sample room
- Optional: UV lamp for decontamination between two runs by 254 nm UV light
|
Kits and Reagents | Ready-to-use kits for DNA and RNA extraction |
Additional | Priming Station and Sample Tray |
Interface | Tablet: USB for data transfer, barcode reader |
Noise emission | Max. 55 dB |