- RapidVap N2 and N2/48 Dry Evaporation Systems combine vortex motion, dry heat and nitrogen blow down for fast evaporation rates
- N2 models for moderate sample throughput with end-point determination include an 8-place sample block, introduces nitrogen to tubes in sets of two and access up to 450 mL of samples per tube
- N2/48 models for small to moderate sample throughput with a 48 sample capacity introduce nitrogen in clusters of six
- Select models available with an RS-232 Link for remote control of functions from a user-supplied computer
- Required: aluminum sample block (2/48), glassware, nitrogen ( 3.5 CFM/100 LPM minimum flow rate), gas pressure regulator
Laboratory Evaporators & Concentrators
6921-PP-01
6921-PP-01 72283
RapidVap® N2 and N2/48 Dry Evaporation Systems by Labconco
Read more
Manufactured By: Labconco
Warranty: One-Year Manufacturer Warranty
Print
Want a quote quickly ?
Submit a QuickQuote request
Select Product Options
Summary
- RapidVap N2 and N2/48 Dry Evaporation Systems combine vortex motion, dry heat and nitrogen blow down for fast evaporation rates
- N2 models for moderate sample throughput with end-point determination include an 8-place sample block, introduces nitrogen to tubes in sets of two and access up to 450 mL of samples per tube
- N2/48 models for small to moderate sample throughput with a 48 sample capacity introduce nitrogen in clusters of six
- Select models available with an RS-232 Link for remote control of functions from a user-supplied computer
- Required: aluminum sample block (2/48), glassware, nitrogen ( 3.5 CFM/100 LPM minimum flow rate), gas pressure regulator
- Unique Cool-Zone™ in RapidVap N2 insulates samples remaining in glassware
- 1000-watt dry block heating system: speeds evaporation and heater and block move in tandem for efficient heat transfer
- Material: powder-coated aluminum base with acrylic/PVC thermoplastic, chemical-resistant glass lid, Phenol-free gasket, PTFE-coated aluminum chamber (for broad range of chemical compatibility)
- Dual lid clamps hold lid securely in place
- Streams nitrogen or dry gas onto sample surface; nitrogen blow down helps remove solvent during evaporation
- Microprocessor-controlled: programming for vortex speed, heat, time, nitrogen blow down and memory:
- Vortex speed: 0 to 100%; heat from OFF to 100° C; time from 1 to 999 minutes
- Heat from OFF to 100° C
- Time: from 1 to 999 minutes
- Active nitrogen positions of 2, 4, 6 or 8
- Memory: stores 1 to 9 programs
- LCD Display: program number, actual and set point vortex, speed, actual and set point temperature, time and number of active nitrogen position (N2 models) number of active nitrogen clusters (N2/48 models)
- Memory storage: 1 to 9
- Control Panel: RUN/STOP button, PREHEAT/END ALARM button and indicator light, set point selection buttons, and increase/decrease programming buttons
- Brushless DC motor: belt-driven, non-sparking, smooth variable-speed power drives vortex motion up to 500 rpm
- Alarms: audible alarm with automatic shut down for timed end point, audible/visual alarm for end point detected by temperature sensors at heater block
- Mechanical components isolated from chemical fumes and vapors
- Nitrogen inlet hose, 1/4" ID
- Electrical:
- 115V models include 6.5', three-wire electrical cord with 20A plug. Requires 15A minimum circuit breaker
- 230V models include 6.5', three-wire electrical cord with 15A plug. Requires 8A minimum circuit breaker
Other products you may find relevant